Detector Response NonlinearityWhile the response of bands 1-4 is linear over the SOFIE dynamic range, bands 5-16 exhibit some response nonlinearity at high radiances. This nonlinearity was calibrated in the laboratory using the small attenuator technique. The measured response, NM (counts), is corrected byNL = NM / fNL(NM) fNL(NM) = 1 - K NM and K is the nonlinearity
constant. The results from three calibration sesions are summarized in the table
below. The baseline nonlinearity
constants are from the October 2005 calibration, which covered the
full dynamic range due to the absence of the CSM flight neutral density
filter.
|